泰亚赛福 —— 世界领先的检测仪器集成供应商
努力为您创造价值
智能 编号 品牌
 您当前的位置:首页 >> 工业检测仪器设备 >> 气体分析及烟气/尾气监测设备 >> 氧气分析仪 >>Model AT300Model AT300
“氧气分析仪”专销电话
专销电话
电 话:010-84851836
产品分类
ABCDEFGHIJKLMNOPQRSTUVWXYZ
最近访问过的商品

Model AT300Model AT300

添加时间:2010-6-2 16:03:00 担当:未知 阅读:

  • 产品名称:Model AT300Model AT300
  • 定货号:YE-00928-00
  • 推荐:
In situ mounting 
High response time
Temperature up to 1600°C
No calibration for drift needed
Suitable for dusty process
Inexpensive
 
This analyzer based on an yttria stabilized zirconia cell, is an excellent solution for oxygen measuring in the regeneration chambers or in the first part of the smoke flue. The precise compensation of the output of the zirconia cell to the real temperature of the platinum electrodes, measured by the built in thermocouple makes the analyzer to be quite free from any drift both of zero and span. A test has demonstrated stability within 0.1% over 12 months. The probe is mounted inside a protection tube that guarantee a long trouble free working service without any service. The analyzer has been drawn and tested for use in side port , end port and unit melter furnaces.
 
The system is composed by the zirconia probe (produced in two different length) the electronic converter. The analyzer does not need any reference air.
 
The probe mod. 4153 is very compact and resistant it includes a fully yttria stabilized zirconia cell the platinum electrode, the outside ricristalized pure alumina tube and the B type compensation thermocouple. The head is spaced from the threaded nipple to avoid over temperature of the electric connections. The probe must be mounted in vertical position, inside the protection electrofused tube that guarantees a long service and a good response time. The vertical position of the probe enables the convective movements inside the reference compartment of the cell to keep the O2 value at 21.0% without any need for sending to the probe reference air.
 
The converter mod. H705 is a panel transmitter that can convert the electric signals incoming from zirconia cell and the compensation thermocouple type B of the FER high temperature oxygen probes mod. 4153. Microprocessor based connected to an analogue to digital 16 bit converter, is equipped with a high brightness led display (31/2 digit) that continuously visualize the measured oxygen and pressing a key the temperature measured by the compensation thermocouple, the resistance of the cell at high temperature to evaluate the wear of the electrode in contact with the process and all settable parameters. The junction box is equipped with a connector to make the cabling easy.
 

定货信息:
定货号 产品名称 规格配置 价格
YE-00928-00标准配置或电话咨询010-84851836咨询