泰亚赛福 —— 世界领先的检测仪器集成供应商
努力为您创造价值
智能 编号 品牌
 您当前的位置:首页 >> 工业检测仪器设备 >> 气体检测仪 >> 固定式有毒/氧气检测仪 >>SensAlert -- 121042-D-1SensAlert -- 121042-D-1三氢砷化产品
“固定式有毒/氧气检测仪”专销电话
专销电话
电 话:010-84851836
产品分类
ABCDEFGHIJKLMNOPQRSTUVWXYZ
最近访问过的商品

SensAlert -- 121042-D-1SensAlert -- 121042-D-1三氢砷化产品

添加时间:2010-6-4 19:24:35 担当:未知 阅读:

  • 产品名称:SensAlert -- 121042-D-1SensAlert -- 121042-D-1三氢砷化产品
  • 定货号:YD-01887-00
  • 推荐:
Minimum Indicated Concentration .............. 0.04 ppm
Repeatability .............................................. ± 2% of reading
Accuracy * ................................................. ± 2% of full scale
Zero Drift ................................................... < 5% change per year (typical)
Span Drift .................................................. < 10% change per year (typical)
Response Time (Rise) ................................. T50: < 10 seconds, (typical)
T90: < 30 seconds, successive exposures
Recovery Time (Fall) ................................... T10: < 60 seconds
Temperature Range ................................... -20° to 50°C (-4° to 122°F)
Humidity Range (continuous) ...................... 5–95 %RH, non-condensing
Humidity Range (intermittent†) ................... 0–99 %RH, non-condensing
Pressure Range .......................................... Ambient atmospheric, ± 1 psi
Recommended Calibration Flow rate .......... 1.0 LPM
Oxygen Requirement .................................. 1% by volume, minimum
† Gas exposure should not exceed eight (8) hours during any 24 hour period.
* When unit is calibrated and serviced at recommended intervals.
ARSINE
Interferent TLV LEL Exposure Response
Acetylene asphyxiant 2.5 %v ppm-range No Data-Pos(+) Interferent
Carbon Monoxide 25 ppm 12.5 %v 300 ppm None
Diborane 0.1 ppm 0.8 %v 1.3 ppm + 1 ppm
Germane 0.2 ppm *** 0.5 ppm + 1 ppm
Hydrocarbons %vol range None
Hydrogen asphyxiant 4.0 %v 3000 ppm + 1 ppm
Phosphine 0.3 ppm *** 0.8 ppm + 1 ppm
Silane 5 ppm *** 1.8 ppm + 1 ppm
Sulfur Dioxide 2 ppm *** 4 ppm No DataMinimum Indicated Concentration .............. 0.04 ppm
Repeatability .............................................. ± 2% of reading
Accuracy * ................................................. ± 2% of full scale
Zero Drift ................................................... < 5% change per year (typical)
Span Drift .................................................. < 10% change per year (typical)
Response Time (Rise) ................................. T50: < 10 seconds, (typical)
T90: < 30 seconds, successive exposures
Recovery Time (Fall) ................................... T10: < 60 seconds
Temperature Range ................................... -20° to 50°C (-4° to 122°F)
Humidity Range (continuous) ...................... 5–95 %RH, non-condensing
Humidity Range (intermittent†) ................... 0–99 %RH, non-condensing
Pressure Range .......................................... Ambient atmospheric, ± 1 psi
Recommended Calibration Flow rate .......... 1.0 LPM
Oxygen Requirement .................................. 1% by volume, minimum
† Gas exposure should not exceed eight (8) hours during any 24 hour period.
* When unit is calibrated and serviced at recommended intervals.
ARSINE
Interferent TLV LEL Exposure Response
Acetylene asphyxiant 2.5 %v ppm-range No Data-Pos(+) Interferent
Carbon Monoxide 25 ppm 12.5 %v 300 ppm None
Diborane 0.1 ppm 0.8 %v 1.3 ppm + 1 ppm
Germane 0.2 ppm *** 0.5 ppm + 1 ppm
Hydrocarbons %vol range None
Hydrogen asphyxiant 4.0 %v 3000 ppm + 1 ppm
Phosphine 0.3 ppm *** 0.8 ppm + 1 ppm
Silane 5 ppm *** 1.8 ppm + 1 ppm
Sulfur Dioxide 2 ppm *** 4 ppm No Data

定货信息:
定货号 产品名称 规格配置 价格
YD-01887-00标准配置或电话咨询010-84851836咨询
相关产品