UD2V-P45.Lite超声波探伤仪
        添加时间:2010-3-31  编辑:C055  阅读:次
	
	
        
        
        
        
          
            - 产品名称:UD2V-P45.Lite超声波探伤仪
- 产品定货号:MC-07548-00
 
        
       
      
        New digital ultrasonic flaw detector, created for replacement of out-of-date analog devices UD2-12. Keeping opportunities of analog processing and having completely a digital path the device allows carrying out all functions ultrasonic flaw detector UD2V-P45/46 except for preservation of results of the testing and communication with the personal computer. Ultrasonic flaw detector can be modernized in the further up to version UD2V-P46.
It is intended for detection of defects (discontinuity and uniformity of materials) in semi-finished items, finished articles and welded joints, for measurement of depth and their coordinates, measurements of thickness, measurement of velocity of propagation and attenuation of ultrasonic oscillations in a material.
Has the high-contrast liquid crystal display, it is simple and convenient in circulation.
 
| Range: 3 to 2975 mm  Display: LCD 110 mm x 65 mm (240 x 128 pixels) with switchable backlight  Controllable gain: 110 dB  TGC/DAC function  Alarms for gates  Pulse echo and dual operation  PRF: up to 800 Hz to 40Hz  Rectification: RF (in all ranges), full-wave, positive and negative half wave  Reject: 0 to 80% screen height  | Two independent monitor gates with individual logic  A-scan image freeze and zoom  Trigonometric functions  Digital measurement resolution: 0.001 mm up to 10 mm steel  Memory: 12 datasets with A-scan and no measurements results  Mains power: 4 NiMh batteries size "D" or via external plug in power unit  16 hours operating with backlight  Weight: 2.5 kg with batteries  | 
 
Basic Delivery Set
UD2V-P45.Lite main unit 
AC-DC power supply 220V 
Probe cable CR 50-Lemo 2 pieces 
Two transducers 
Batteries set unit NiMh size "D" 9000 mA/hr 
Documentations 
Transport bag 
Additional implements
Ultrasonic transducers for differently application 
Standard samples 
Тechnical Characteristics 
Calibration range 
min.: 0 - 2.9 mm (steel); 0 - 1 mks
max.: 0 - 2975 mm (steel); 0 - 1000 mks 
Sound velocity range: 
1000 - 9999 m/s 
Pulse shift 
up to -0,5 mks to 996 mks 
Probe delay: 
0 - 100 mks 
Damping 
50 ohms / 1000 ohms 
Resiver 
50 ohms / 600 ohms 
Pulse intensity 
quadrangular, amplitude 200V,
with change interval up to 50 to 500 ns,
with step 25 ns 
Pulse repetition frequency 
up to 800 Hz or 40 Hz 
Amplifier 
broadband 1-10 MHz (-3 dB) 
Controllable gain 
110 dB, with step 0.5, 1, 2 or 6 dB 
TGC/DAC 
curve recording with maximum of 10 reference echoes. 
Rectification 
RF (in all ranges), full-wave, 
positive and negative half wave 
Reject 
linear, 0-80% screen height 
Monitor gate 
two independent monitor gates with individual logic, variable in steps of 1% 
Alarm 
sonorous and luminous 
Sound path measurement 
digital display of sound path between zero and the first echo signal
in the gate or between two echoes in the gates, measurement at echo flank
or peak with trigonometric flaw location. 
Measurement resolution 
3.125, 6.25, 12.5 or 25 ns depending on PRF 
Display of amplitude 
as a percentage of the screen height
in dB to reference test echo
in dB reference to DAC curve. 
Display 
LCD with switchable backlight and 
(110 mm x 65 mm; 240 x 128 pixels) 
А-signal 
73 mm x 50 mm; 160 x 100 pixels 
Memory 
12 datasets with А-signal 
Interface with PC 
no 
Dimensions transducers 
2 SR50 
Batteries 
4 NiMh size "C" or "D" 
Operating time 
16 hours from batteries 9 А/hr "D"
with backlight
Power supply 
external block 220 V AC 
Operating voltage 
4 - 10 V DC 
Power consumption 
maximum 5 W, depending on instrument setting 
Operating temperature 
up to +5 C to +55 C 
Dimensions (H x W x D) 
160 mm x 225 mm x 45 mm 
Weight 
2,5 kg with batteries