Housing: |
Stainless steel with 40 mm insulation, (material 1.4301/ SS304) / PS / PC (GF10) Option: material 1.4571/ SS316Ti; SS304 EPOXY-coated; SS316Ti EPOXY-coated |
Thermostatic control: |
Self-contained cooling / heating with 4 settings, no-frost. Temperature in sample compartment: 4°C (adjustable from 0,0-9,9°C) |
Control: |
Microprocessor control, foil keyboard, back lit four-line LC-display |
Data logger: |
Nonvolatile data memory; storage of sampling and malfunction data like sample extractions, bottle changes, messages, external signals |
Programming: |
6 user programs (can be edited freely) |
Program start options: |
Immediately; at a certain time; by an external signal |
Program stop options: |
End of sampling program after one program run; continuous operation |
Pause mode: |
Interruption of program run at any time |
Overfilling protection: |
Adjustable from 1–999 samples/bottle |
Interval setting: |
1 min. to 99 h 59 min. in steps of 1 minute |
Pulse setting: |
1 to 255 pulses/sample |
Manual sample extraction: |
Possible at any time without interrupting the current program run |
Program protection: |
Up to 5 years after voltage loss |
Interface: |
RS 232 |
Communication: |
Modbus, connection via DP PROFIBUS optional |
Modem GSM communication: |
Optional (in combination with PC software) |
Languages: |
Multi-language, selectable |
Signal inputs: |
• 1 x analogue: 4-20 mA, optional 0-20 mA, minimum voltage 3,3 V (optocoupler) • 5 x digital (flow, event, 3 inputs can be programmed freely) |
Signal outputs status messages: |
Max. 8 digital outputs, depending on device version; freely programmable |
Sampling method: |
-Vacuum system 20-350 ml -Vacuum system 20-500 ml optional -Vacuum flow-proportional system 20-350 ml optional -Bypass system 20-250 ml optional |
Single sample volume accuracy: |
< 2,8 % |
Suction height: |
max. 8 m (at 1013hPa and stagnant medium) |
Pumping speed: |
>0,5 m/s at suction height up to at least 7,8 m (at 1013h Pa); pump capacity can be adjusted electronically |
Suction hose: |
PVC, L=7,5 m, ID=12 mm (ID 16 mm at flow-proportional option). Max. hose length 30 m |
Sampling modes: |
Time-related, flow-dependent, flow-proportional, event-related, manual sample extraction |
Bottle variants: |
Plastic 1 x 25 l, 1 x 50 l, 2 x 10 l 4 x 6,0 l, 4 x 10 l, 4 x 14 l, 12 x 2,9 l, 24 x 1,0 l Glass 12 x 2,0 l 24 x 1,0 l |
Overall dimensions: |
Standard (hxwxd): 1.290 (1.890*) x 690 x 645 mm As measuring station: 1.470 (2.070*)x 690 x 645 mm *) with opened top |
Weight: |
100 kg with composite container, higher weight when using several bottles and/or glass bottles |
Power supply: |
230 V / 115 V /AC |
Power requirement number of samples: |
Approx. 350VA (with cooling) |
Ambient temperature: |
-20 – 43°C |
Sample temperature: |
0 – 40°C |
Standards: |
Device meets ISO 5667 standard |
Wetted materials: |
PVC, Silicone, PS, PE, EPDM (optional: metering vessel Duran50, sinker weight SS304) |
Accessoires: |
Plinth Extraction unit, fixed or pendulous Mobile version Strainer basket PC communication software Status messages Interior lighting Main switch
|