The Advanced Instruments Model 2020 Micro Osmometer determines the osmolality of solutions using freezing point depression (FPD). It is ideal for serum, urine or other bodily fluids in clinical applications; and media or formulations in pharmaceutical, biotech or industrial applications. The Model 2020 can process up to 20 samples at one time.
Automated, multi-sample, designed to process a 20uL sample with a 90-second test time. The Model 2020 provides maximum productivity, accuracy and reliability.
• Up to 20 samples at one time
• 20 uL sample size is ideal for applications where the sample quantity is limited.
• Stat sample capability
• Multi-language operation
• On-board statistical analysis
• Results recall up to the last 200 tests
• On-board printer
• Connects easily to external barcode scanner
• Internal diagnostics
• Ergonomic, compact design
• Automated calibration
• Hot-Line® Technical Service available 24/7
Advanced Instruments also supplies ControLine Products
• ClinitrolTM 290 mOsmol Reference Solution
• Protinol® Protein-based Controls
• RenolTM Urine Osmolality Contols
• Linearity Osmolality Controls
Method – Freezing Point Depression Osmometry
Osmometry measures the total molar concentration of dissolved solids in any solution. The freezing point of a solution is depressed in direct relation to the amount of solute in solution.
Uses and Applications – Emergency Medicine, Internal Medicine/Endocrinology, Urology/Renal Function, Surgery/Critical Care, Obstetrics/Gynecology, Pharmacy, General Quality Control.
Advanced® Model 2020 Osmometer Technical Specifications
Electrical
Power Requirement: 100 to 240 VAC (50-60 Hz)
Fuses (2): 250V time delay (Type T): 2-Amp
Power Consumption: 80 Watts
Memory Backup: Integral lithium cell; 10-year life (typical); (not user-replaceable)
Sample Volume: 20 μL
Sample Capacity: Up to 20 samples
Readout: 2 line by 24-character vacuum fluorescent display
Units: mOsm/kg H2O
Range: 0 to 2000 mOsm/kg H2O
Resolution: 1 mOsm/kg H2O
Communications: On-board printer, DTE EIA-232/V.24 (RS-232) serial port and optional barcode scanner
Performance at Reference Conditions1
Linearity: ±3 mOsm/kg between 0 and 300 mOsm/kg H2O; less than ±1% above 300 mOsm/kg H2O
Repeatability: 0 to 400 mOsm: ±3 mOsm/kg (1 S.D.) 400 to 2000 mOsm: ±0.75% (1 S.D.)
Drift: Less than 1 mOsm/kg H2O per month
Performance Over Operating Conditions
Temperature Effects: Less than 1 mOsm/kg H2O for every 5°C (9°F) ambient temperature change
Operating Conditions Temperature: 18°C to 35°C (64°F to 95°F)
Humidity: 30 to 80% relative humidity (non-condensing)
Storage Temp.: -40°C to +45°C (-40°F to +113°F)
Start-up Time: Immediately from stand-by; under 5 minutes from power-on
Test Time: 90 seconds per sample, typical
Sample Throughput: 35 to 40 per hour
Dimensions
29.0cm x 40.6 cm x 31.0 cm
(11.4" H x 16.0" W x 12.2" D)
Weight
Net: 10.0 kg (22.0 lbs.)
Shipping: 15.5 kg (34.0 lbs.)
Warranty: One-year limited warranty on workmanship and all parts except glass, plastic, and parts warranted by their makers