Finally, a portable Dew Point hygrometer that integrates the
accuracy and reliability of the EdgeTech chilled mirror, with the
battery life and ruggedness required for field use.
Introducing the Model 1500 portable chilled mirror hygrometer.
This device incorporates all of the standard features that you’ve
become accustomed to from EdgeTech, such as drift-free NIST
traceable accuracy, low maintenance, high durability and ease
of use, in a system that is ideally suited for the demands of field
use. Whether you use the Model 1500 as a field transfer
standard, or to spot check your process air dryness or Endo gas
dew point, you’ll find this system trouble-free to operate.
The Model 1500 is a self contained system; not only is there
no field assembly required, there’s no need to carry around a
myriad of individual components such as control boxes, sensors,
cables, pumps, and power cords. Simply attach the sample line,
adjust the flow (with the integral flow meter) and turn on the
power. Within minutes you’ll be collecting field dew point data
with the same chilled mirror accuracy of a lab standard
hygrometer. And with a nominal battery life of 12 hours, you will
be able to collect for the entire shift or spot check your processes
for several days.
If you are looking for accurate, flexible, easy to use portable
chilled mirror dew point measurement. Or, if you are tired of
carting around a slew of gear just to make your process dew
point monitor portable, contact your local EdgeTech
representative or call us toll free at 800-276-3729.
You can also visit our web site at ww.edgetech.com;
email at h2o@edgetech.com.
Portable Dew Point Monitor
Mooddeell 11550000
A Generation of Advanced
Optical Chilled Mirror
Sensors
Features
► Primary measurement
technique
► NIST Traceable
► Rechargeable Batteries
► Easily transportable in hard
protective case
► Wide measurement range
(-75 to 75°C)
► Microprocessor controlled
► Self balancing
Model 1500
Toll Free: 800 - 276-3729 Ph: 508 - 263-5900 Fax: 508 - 486-9348
Email: h2o@edgetech.com Web: www.edgetech.com
RANGE:
Dew/Frost Point: Single-Stage (S1 & DS1): -4 to 122°F
-20 to 50°C
Two-Stage (S2 & DS2): -40 to 122°F
-40 to 50°C
Three-Stage (S3): -103 to 122°F
-75 to 50°C
S3 sensors are not battery powered
ACCURACY:
Dew / Frost Point: +/- .36°F (.20°C)
Dimensions: 14.75” H; 18” W; 7”D
Weight: 24lbs.
Enclosure: Plastic carrying case
► Temperature Probe (PRT) w/ %RH
► Live pressure w/ PPMv (S-type sensor only)
► D-Type sensors (probe style) available
► RS232
► Internal Rotary Vane Vacuum Pump
M E A S U R M E N T S P H Y S I C A L
O P T I O N A L F E A T U R E S
Power/Charger: 115-230 VAC, 50-600Hz
Operating Temperature: 32 to 122°F (0 to 50°C)
Sample Connection: 1/4” Compression Fitting
Sample Flow Rate: 0.5 - 5 SCFH; Integral Flow Meter
Sample Pressure: 0 - 150 p.s.i.g (higher pressures
available - consult factory)
Outputs:
Analog: 4 to 20mA (or 0-5VDC), -50 to 50°C
or -75 to 75°C
Compliance: 9.0VDC, 450 ohms
Display: Red LED, 0.5” tall;
8 digit alphanumeric
Alarm:
Visual: Flashing “ALARM” message
Relay: 1 Form C, NON-latching
10A @240 VAC, 8A @24 VDC,
1/2 HP @ 240 VAC
F U N C T I O N A L I T Y
► Programmable Automatic Balance Control (ABC)
► 8-digit Alphanumeric Display
► User Programmable Alarm
► Internal Flow Meter and Valve (S-type only)
► Visual Alarms
► 4 - 20 mA (or 0-5VDC), Analog Output
► Microprocessor Controlled
► 115/230 VAC Battery Charger