Hydrogen Sensitivity Range: 0.5% to 100% hydrogen by volume at 1ATM Typical Response Time: T90 less than 30 sec Ingress Protection: IP64 capable Calibration Interval: 90 days Product Life Expectancy: 10 years Accuracy: ± (0.03 x indication + 0.2) percent hydrogen by volume |
Input Voltage Range: 8 VDC to 13 VDC Input Current: 0.5 A Analog Outputs: Output Voltages - OR - Output Currents • 0 VDC to 5 VDC • 4 mA to 20 mA • 1 VDC to 5 VDC • 0 mA to 20 mA • 0 VDC to 4 VDC • User-specific mA range • 0.5 VDC to 4.5 VDC • User-specific VDC range Serial Communication Options: RS232 or RS422 Relay Contacts: Two programmable 60 VDC / 1 A SPDT relays with both normally open (N.O.) and normally closed (N.C.) contacts. |
½ in. MNPT thread ½ in. FNPT thread -8 SAE/MS thread size ... and many other industry standards |
Operating Humidity: 0% to 95% RH Flow Rate: 0.1 slpm to 50 slpm Pressure: 0 to 7 bar gauge, 0 to 100 psi gauge Process Gas Temperature Range: -20ºC to 100ºC Operating Temperature Range: -20ºC to 40ºC Storage Temperature Range: -40ºC to 50ºC Calibration Background Gas: Nitrogen |
定货号 | 产品名称 | 规格配置 / 简介 | 市场价/(会员价) |
---|---|---|---|
MM-38373-00 | HY-OPTIMA™ 700氢气监控仪 | /() |