Hand-held measurement of relevant climatic data is becoming
more and more important. Thermo-Hygrometers of the
HUMIPORT series set new standards because of the
combination of the latest technology and design.
Excellent readability of the large and illuminated display, the
simple thumbwheel operation and the robust case allow the use
in tough industrial environments. Latest technology and the
well proven E+E humidity sensor guarantee highest accuracy
and long term stable measurements.
The HUMIPORT measures relative humidity [% RH] and
temperature [°C or °F] and calculates dew point
temperature [°C or °F], absolute humidity [g/m³] and mixing
ratio [g/kg or gr/lb].
HUMIPORT10/20 Thermo-Hygrometers are available in a
practical carrying case. The set contains a HUMIPORT incl. an
E+E factory certificate and a humidity calibration device with 5
ampoules calibration fluid (80% RH).
Thermo - hygrometer and laser pyrometer of the HUMIPORT 05 IR
series provide 3 operation modes:
TH-mode: - Temperature / humidity measurement
- In this mode the hand-held offers you all
functions of the HUMIPORT 05 series.
IR-mode, infrared: - Contact-free surface temperature measurement
with measurement position marking
DP-mode: - Dew point temperature / surface temperature
HUMIPORT 05 HUMIPORT 05 IR HUMIPORT 10/20
Measuring values
Sensor type relative humidity HC101 HC101 HC101
temperature NTC NTC NTC
surface temperature Thermopile
Measuring relative humidity 5...95% RH 5...95% RH 0...98% RH
range temperature -20...50°C (-4...122°F) -20...50°C (-4...122°F) -20...50°C (-4...122°F)
dew point at 20°C (68°F) -15...19°C (5...66°F) -20...19°C (-4...66°F)
mixing ratio at 20°C (68°F) 1.18...14g/kg (8.26...98gr/lb) 0.78...14g/kg (5.46...98gr/lb)
surface temperature -20...60°C (-4...140°F)
Accuracy relative humidity ±2% RH ±2% RH ±2% RH
temperature ±0.4°C (0...40°C) ±0.4°C (0...40°C) ±0.3°C (0...40°C)
(±0.7°F at 32...104°F) (±0.7°F at 32...104°F) (±0.54°F at 32...104°F)
±0.7°C (-20...50°C) ±0.7°C (-20...50°C) ±0.5°C (-20...50°C)
(±1.3°F at -4...122°F) (±1.3°F at -4...122°F) (±0.9°F at -4...122°F)
dew point at 20°C (68°F) ± (4.46...0.64°C) ± (4.33...0.45°C)
± (8.03...1.15°F) ± (7.79...0.81°F)
mixing ratio at 20°C (68°F) ± (0.44...0.57g/kg) ± (0.29...0.4g/kg)
± (3.08...3.99gr/lb) ± (2.03...2.8gr/lb)
surface temperature ±2°C (±3.6°F)
(Tobj > 0°C (32°F);
Tamb > 10°C (50°F))
Resolution relative humidity 0.1% RH 0.1% RH 0.1% RH
temperature 0.1°C (0.18°F) 0.1°C (0.18°F) 0.1°C (0.18°F)
surface temperature 0.1°C (0.18°F)
General
Supply voltage 9V battery 9V battery 4x 1.5V AA battery
Battery lifetime typ. > 150h max. 200h - TH mode typ. > 500h
10h - IR/DP mode
Working temperature range housing: 0...50°C housing: 0...50°C housing: 0...50°C
(32...122°F) (32...122°F) (32...122°F)
probe: -20...50°C probe: -20...50°C probe: -20...50°C
(-4...122°F) (-4...122°F) (-4...122°F)
Storage temperature range -30...60°C (-22...140°F) -30...60°C (-22...140°F) -30...60°C (-22...140°F)
Display LCD, 45x32mm LCD, 45x32mm (1.8x1.3”) LCD, 90x50mm (3.5x1.9”)
(1.8x1.3”) illuminated
Measuring optics1) ~ 8:1
CE compatibility according EN61000-6-2 EN61000-6-2 EN50081-2 EN50082-2
EN50147-3 EN50147-3 EN61000-4-2 EN61000-4-3
EN55011
Housing / protection class ABS / IP40 ABS / IP40 ABS / IP40
Dimensions (HxWxD) housing: housing: housing:
140 x 48 x 25mm 140 x 48 x 25mm 145 x 85 x 37mm
(5.5.x1.9x1”) (5.5.x1.9x1”) (5.7x3.3x1.5”)
HUMIPORT 20: cable
length: approx. 1.5m (4.9ft)
Weight ca. 200g (0.5 lbs) ca. 200g (0.5 lbs) ca. 400g (1 lbs)