The analyzer is based on a yttria stabilized zirconia cell, that supply the relevant advantages that this measuring principle can guarantee. The absence of cross sensitivity with other gases due to the oxygen vacancy generated in the zirconia substrate is one of the most important feature. 
 
The peculiar drawing of the analysis chamber allows the changing of the flow of the sample without affecting the O2 measure. 
 
The analyzer does not require regular calibrations because it is fully temperature compensated. The measuring principle that does not need any moving parts, ensures a very long life without any service. 
 
The measuring range by volume is very wide O2 values and diagnostic are transmitted with analogical, serial and digital output. Optionally the Hart protocol is available. 
 
This very tough instrument, is ideal to be used in a system for emission monitoring. It cannot be used for measuring O2 in a gas that contains combustibles, because the zirconia cell is warmed up to high temperature. 
 
TECHNICAL SPECIFICATIONS
 
Gas analyzed:  Oxygen in gas phase 
Measuring principle:  Stabilized zirconia 
Measuring field: 1 ppm – 25% O2 by volume 
Accuracy:  +/- 0.5 of true value or 0.1% O2 whichever is the greater 
Repeatability:  Within  1% 
Response time  (90%):  < 10 sec. With a flow of 3 l/min 
Flow of the sample gas:  0.5 – 3 l/min 
Flow adjusting:  Niddle valve of the flowmeter. 
Analog outputs:  4 – 20 mA isolated, max load 500 ohm or 0 – 10 V  10 mA linear on one of the following selectable fields: 
0 – 1999 ppm;  
0 – 5%;  
0 – 10%;  
0 – 25% , if the value is not valid the output is switched to 2 mA. 
Serial outputs:  Mod Bus RTU su RS 485
Options:  Hart protocol 
Alarms:  Potential free for: O2 high, O2 low, measure not valid, servicing, lack of flow. For all  250 V, 1 A max. 
Dumper:  Time constant selectable from 0 and 900 sec. 
Power supply:  230 e 115 V AC +/- 10%  50/60 Hz  300 VA max 
EMC:  According EN 50081 e 50082 
Electrical safety:  According EN 61010 – 1 
Temperature limits:  0 – 40°C 
Humidity:  < 90% non condensating 
Storage temperature:  - 40 / +80°C 
Sample gas input and output:  tubing  mm. 6x4 
Test gas input:  tubing  mm. 6x4 
Housing:  Rack 19” 3 units  IP20 
Dimensions:  450x132 mm, deep 380 mm 
Weight:  9 Kg approx.