The Series 2520 Portable Percent Oxygen Analyzer is a light-weight, easy-to-use analyzer that provides accurate and repeatable percent oxygen measurements in a variety of gases. The rugged and compact design of the Series 2520 makes it ideal for industrial applications where spot oxygen measurements need to be made.
System Description
The Series 2520 Portable Percent Oxygen Analyzer is a light-weight, easy-to-use analyzer that provides accurate and repeatable percent oxygen measurements in a variety of gases. The rugged and compact design of the Series 2520 makes it ideal for industrial applications where spot oxygen measurements need to be made. The Series 2520 can be configured with one of several measuring ranges from 0-1% to 0-100% oxygen. The standard Series 2520 Portable Percent Oxygen Analyzer is equipped with a 3-1/2 digit liquid crystal display, in-line filter, built-in NICAD batteries, and an AC adapter that provides the ability to recharge the NICAD batteries even while powering the analyzer from the adapter.
Specifications
Measurement Ranges: 0-1, 0-5, 0-10, 0-25, 0-50, and 0-100%
Accuracy: ±1% of full scale
Linearity: ±1% of full scale
Response Time: 90% of full scale in <20 seconds
Sensor Type: Electrochemical
Temperature Compensation: Standard
OperationTemperature Range: 40° to 100°F (5° to 38°C)
Sample Gas Pressure Limits: 0.1 to 1.0 psig. (1.04 kg/cm² to 1.10 kg/cm²)
Sample Flow Rate: 1.0 to 2.0 SCFH (0.5 to 1.0 liters/minute)
Power Source: Rechargeable NICAD batteries or 115 VAC/60 Hz using the AC adapter
Analog Output: Not available
Warranty: Two years electronics/1 year sensor
Enclosure: Polycarbonate-rated NEMA 1
Display: 3-1/2 digit liquid crystal display
Gas Connections: Quick connect
Dimensions: 5.8” H x 6.7” W x 3.4” D (dimensions change based on options and accessories)