Coercimeter it is intended for non-destructive testing assurance thermal, thermomechanical or thermochemical processing, and as definitions of hardness and mechanical properties of details from ferromagnetic materials in the presence of correlation communication between supervised and measured in the parameters. The device can be used for grading of metals on marks of steel and the testing of blankets of ferromagnetic materials.
The transducers is an added electromagnet with demountable polar tips and with the gauge of the Holla in its magnetic chain. The principle of work of the device consists in magnetization of a controllable site of a detail with the subsequent demagnetization its accruing field, the fixing of intensity of a field corresponding coercimeter to force of a material of a detail, and measurement of amplitude of a signal from the Holla gauge.
Continuous measurement of residual magnetizations
Adjustable quantity of pulses of magnetisation
Automatic sound alarm signal
Selective averaging up to 99 results |
Memory for storage of 5000 measurement results
10 additional acales programmable by PC or direct from keypad
Powered by 220V plug-in power supply or Li-Ion battery
Weight: 1.2 kg with batteries |
Basic Delivery Set
KIM-2M main unit
Transducer
Power supply 220V/10V
Interface cable RS232 for PC connection
Software CD-ROM
Protective cover
Transport bag
Additional implements
Demountable tips for the transducer
Technical Characteristics coercimeter
Base-time measurement values
coercive force,
residual magnetic induction at partial demagnetization,
residual magnetization
Corcive force measurement range
up to 150 А/m to 4500 А/m
Magnetization
by 220 V pulse; pulses quantity can adjust from 0 to 10
Demagnetization
by preset current 0-1000 mA in steps of 1 mA
Measurement time
5 sec. at 3 magnetization pulses
Scales
15 programmable direct from keypad
or by PC scales, up to 10 points
for linear approximation
Indication accuracy
1, 0.1, 0.01 or 0.001 - is
set at programming scales
Mains power
internal Li-Ion battery or
power unit 220 V AC
Operating time
not less than 5 hours from batteries "АА"
Operating voltage
9 - 12 V
Operating temperature
0 C to +50 C
Dimensions of the electronic block (H x W x D)
205 mm x 115 mm x 50 mm
Dimensions Transducer (H х W х D)
60 mm x 75 mm x 35 mm
Weight with elements batteries
1.2 kg