New model ultrasonic flaw detector, combining advantages of device UD2V-P45 with the expanded functions to measure and record results of the testing:
simultaneous measurement of amplitude and coordinates position of a maximum signal;
simultaneous measurement of distance on-beam and defect coordinates for angle trasdusers;
a mode of record bending around a signal maximum in a examination area, with B-scan display;
the expanded memory options and results of the testing;
software support for several databases of the testing reports on the personal computer with ReportConstructor function
full-screen an operating mode (LD-version)
possibility to connect one of 7 built-in matching elements, for optimum work with any transducers
The algorithm of record realized in the device bending around a signal maximum differs from similar modes in similar devices of other firms. As a rule, it is a so-called mode "Peak freezing", when the device memory the maximum value of amplitude of a signal in each point of range. Unlike such mode, realized in UD2V-P46 the method allows to construct bending around remembering only on one point in each probing impulse - by position and value of the maximum amplitude of an echo-signal in a testing zone.
Such mode gives the chance to receive, for example, the move diagramme maximum a signal at cross-section scanning of defect in welded joints and to estimate its conditional extent, and also it is easy to define reflector coordinates at the maximum amplitude of the reflected signal.
One more considerable improvement in new model - increased to 100 quantity of options, with record of the A-signal and possibility to input a name of adjustment from the device keyboard, and as possibility of preservation to 750 results of the testing (15 files on 50 results). Each result represents the full report of the testing and consists of an A-signal, bending around a maximum (if the result is kept in this mode), digital value of result of measurement (amplitude and coordinates, distance on a beam and coordinates or speed ultrasonic), all parameters of work of the device, date, time and a name of the result entered from the keyboard of the device. All results can be looked through and renamed further both in the device, and by means of the software on the computer. Software UdReport, allowing completely automating processing results of the testing is for this purpose intended, dividing them on different bases of reports and defining various additional parameters of the testing on the basis of the identifiers containing in a name of result. Besides, the program has function of the designer of the reports, allowing creating printing forms of reports of any kind depending on requirements of the concrete user.
Range: up to 2.9 to 2975 mm (steel)
Display: LCD or ELD 110 x 65 mm (240 х 128 pixels) with adjustable backlight and contrast
Amplifier: broadband 0.5 - 15 MHz, the filtres 1.25, 2.5, 5 and 10 MHz
7 contours of the coordination with face values: 0,66; 1; 2,2; 3,3; 4,7; 6,8 and 15 mkHz
Contollable gain: 110 dB with step 0.5, 1, 2 and 6 dB
TGC/DAC 90 dB, 10 dB/mks
Work with curve aplitude-interval: measurement in dB signal relative curve and operation LED
Probing impulse: rectangular 200 V, with adjustable duration
Pulse repetition frequency: to 800 Hz depending on parameters setting/40 Hz |
Rectification: positive halfwave, negative halfwave, full-wave, radiosignal (in all ranges)
Reject: linear, up to 0 to 80% screen height
Monitor gate: two independent monitor gates with individual logic, sound indication
"Freeze" A-signal
Measurement of a thickness on one signal or between two signals
Accuracy of indication of a thickness: 0.001 at thickness to 10 mm
Memory datasets options: 100 options with the A-signal and possibility of input of the name of adjustment from keyboard
Memory results: 750 (15 files on 50) the reports of the testing consisting of the A-signal, bending around, result of measurement, date, time and the name of report
Interface RS232: for transfer of A-range, parameters of work and results of measurements on the personal computer
Mains power: 4 NiMh batteries size "D", external block
maximal 8 hours work from batteries
Weight: 2.5 kg with batteries |
Basic Delivery Set
UD2V-P46 main unit
Power supply 220V/9V
Probe cable SR50-Lemo 2 pieces
Four transducers. (P111-2.5-К12, P111-5-Кb, P121-2.5-40, P121-5-65)
Cable RS232 for connection PC
Software support for PC
Documentations
Transport bag
Additional implements
Protective bag
Module for batteries size "D"
Set batteries NiMh size "D" 9000 mA/hr
Universal charging device
Adapter circuit for connection to USB to port of the PC
Ultrasonic transducers for diverse application
Standard samples
Technical Characteristics
Calibration Range
min.: 0 - 2.9 mm (steel); 0 - 1 mks
max.: 0 - 2975 mm (steel); 0 - 1000 mks
Sound velocity range
1000 - 9999 m/s
Pulse shift
up to -0,5 mks to 996 mks
Pulse delay
0 - 100 mks
Damping
50 ohms / 1000 ohms (to 25 ohms in simultaneos modes)
Input impedance
50 ohms / 600 ohms
Soud impulse
quadrangular, amplitude 200 V,
with change interval up to 50 to 500 ns,
with step 25 ns
Pulse repetition frequency
to 800 Hz in a mode of the maximum frequency and 40 Hz
Amplifier
broadband 0.5-15 MHz (-6 dB)
the filtres 1.25, 2.5, 5 and 10 MHz
The coordination with PET foreign manufacturers
7 contours of the coordination with face values:
0,66; 1; 2,2; 3,3; 4,7; 6,8 and 15 mkHz
Contollable gain
110 dB, with step 0.5, 1, 2 or 6 dB
TGC/DAC
90 dB, 10 dB/mks
with curve construction on 10 reference points
entered manually or from testing reflectors
Rectification
positive or negative half wave,
complete, radiofrequency signal
Reject
linear, 0 - 80% screen height
Monitoring gate
two intependent monitor gates with individual logic,
sound indication
Alarm
luminous and sonorous
Operating modes alarm
defect in the first zone,
defect in the second zone,
defect in the first and in the second zones,
defect in one of zones,
on DAC - signal comparison
in the first zone from a curve amplitude-distance.
Sound path measurement
up to 0 to first signal in zone or between signals in zones
Measurement resolution
3.125, 6.25, 12.5 or 25 ns depending
on setting frequency
Display of amplitude
as a procentage of the screen height,
in dB to referencetest echo,
in dB referens to DAC curve
Display
LCD or ELD with adjustable contrast and backlight,
110 х 65 mm; 240 х 128 pixels
А-signal
73 mm x 50 mm; 160 x 100 pixels
in mode "full-screen": 240х128 pixels
Memory
100 datasets with А-signal and name,
750 (15 files on 50) the reports of the testing
Interface
RS232
Dimensions transducer
2 SR50 or Lemo1
Batteries
4 NiMh size "D"
Operating temperature
LCD : 12-16 hours (depending on setting);
ELD: 7-8 hours (depending on setting)
Mains power
via external plug-in power unit 220 V AC
Operating temperature
LCD: up to -5°C to +55°C
ELD: up to -25°С to + 55°С
Dimensions (H x W x D)
160 mm x 225 mm x 45 mm
Weight
2.5 kg with 4 batteries