Advanced Instruments Bilirubin STAT-Analyzer™, Model BR2, for Total Bilirubin and Direct Bilirubin Measurement
 
The Advanced Instruments Model BR2 Bilirubin STAT-Analyzer™ employs the method of Evelyn-Malloy, after Henry, to measure total bilirubin and direct bilirubin in neonatal samples.
 
The BR2 determines bilirubin levels through spectrophotometric analysis.
Works with infant plasma or serum samples
Requires only 30 µL of sample
Total bilirubin measured immediately, direct bilirubin takes just two minutes.
Easy to use, simple setup, calibration and operation
Uses and Applications
Identify hyperbilirubinema
Chart the normal course of post-partum physiological jaundice
Monitor those at high risk of kernicterus
Determine the onset of normal hepatic bilirubin conjugating mechanism
Advanced® Bilirubin Stat-Analyzer™ Photometer Technical Specifications
Electrical
Power Requirement: 100 to 130 VAC (60 Hz) or 200 to 250 VAC (50 Hz)
Fuses (2): 250 VAC time delay (Type T): 1/4-Amp (100-130 VAC), 1/8-Amp (200-250 VAC)
Power Consumption: 15 Watts
STAT Mode: Power left on continuously
Test Capability: Total and Direct Bilirubin, both on the same serum or plasma sample
Sample Volume: 30 μL
Sample Capacity: Single sample, disposable cuvettes
Readout: Three-digit, seven-segment LED display
Units: mg/dL or μmol/L (S.I. units for export only; 1 mg/dL bilirubin = 17.1 μmol/L)
Range: 0 to 30 mg/dL
Linearity: Within ±0.2 mg/dL from zero to 30 mg/dL (straight-line method)
Resolution: ±0.1 mg/dL
Repeatability: Within ± 0.2 mg/dL over entire range
Hemolysis: No interference from HbO2 levels up to 600 mg/dL
Turbidity: Negligible interference from turbidity associated with triglyceride (TGL) levels up to 500 mg/dL
Storage Temperature: -40 to +70°C (-40 to +160°F)
Operating Temp. (ambient): 15 to 32°C (59 to 90°F)
Drift:
Zero: Less than 0.2 mg/dL/hr after 30-minute warm-up
Calibration: Less than 0.2 mg/dL/24 hr after 30-minute warm-up
Room Humidity: 5 to 80% R.H. (non-condensing)
Test Time: Instantaneous for TOTAL Bilirubin, 2 minutes for DIRECT Bilirubin

Dimensions 
18 cm x 28 cm x 33 cm
(7" H x 11" W x 13" D)
Weight
Net:  4 kg (9 lbs.)
Shipping:  5.5 kg (12.0 lbs.)

Warranty: One-year limited warranty on workmanship and all parts except glass, plastic, and parts warranted by their makers.