泰亚赛福 —— 世界领先的检测仪器集成供应商
努力为您创造价值
智能 编号 品牌
通用环境检测仪器
产品导航
 您当前的位置:环境安全 >> 通用环境检测仪器 >> 噪音计 >> Model105&106音位校正器

Model105&106音位校正器

添加时间:2009-12-4 编辑:C055 阅读:

  • 产品名称:Model105&106音位校正器
  • 产品定货号:MB-01815-00
  • 产地:英国
  • 推荐:
Acoustic Calibrators
Models 105 and Model 106 are new technology sound calibrators that have been tested and formally EU Pattern Approved under certificate PTB-1.61.4028829, as meeting all the requirements of the latest International standard, IEC 60942 : 2003.
 
This version of the IEC standard is far more stringent than previous versions and so far very few manufacturers have managed to meet it.
 
The two models are intended to calibrate, that is set the sensitivity, of all Pulsar sound level meters or any properly approved sound level meter that uses a standard measurement microphone.
 
Both Models 105 and Model 106 give a standardised 1kHz tone - the frequency where all weighting networks are the same, at a pressure of 1Pascal which is a sound level of 94 decibel. They have sensors to correct for temperature, humidity, barometric pressure and battery voltage so that they remain stable and accurate under almost any conditions where they are likely to be used.
 
Model 105 complies fully with Class 1 - the precision class, while Model 106 fully meets the requirements for the calibration of industrial sound level meters used in almost all Health and Safety applications. As today, almost all approved sound level meters use the Internationally standardised “1/2inch” microphone, the microphone cavity is this diameter, but adapters are available for smaller microphones such as 1/4 inch.
 

产品名称产品简介产品编号
Model105&106音位校正器MB-01815-00



相关产品