泰亚赛福 —— 工业设备集采平台
努力为您创造价值
智能 编号 品牌
 您当前的位置:首页 >> 品牌 >> SEFRAM >> LCR表
产品目录
品牌简介

  Sefram为B+K Precision集团旗下品牌之一,公司设立于法国,创建于1947年,拥有60年的丰富经验与高度信誉。Sefram专致于记录仪及电视信号分析仪器的生产与研究,产品于欧美拥有极高的声誉,Sefram将以卓越的技术为您提供最精准的测量与分析。

>>品牌详情

推荐产品
热销产品

BK878A通用LCR测试器

2009/12/10 16:12:16 阅读:557次

  • 产品系列名称:BK878A通用LCR测试器
  • 产品定货号:ME-01879-00
  • 产地:法国
  • 品牌:SEFRAM
  • 推荐:
The 878A measures capacitance, resistance (of non-inductive components) and inductance. Components can be measured with selectable test frequencies in series or parallel modes as desired. The 4 1/2 digit main LCD displays values to 19,999, and the secondary 3 digit 1,000 count display reads D or Q. Both models can be used in either auto ranging or manual ranging modes.
 
Features
Selectable test frequencies 120Hz, 1KHz
Simultaneously displays measured component value and Q or Dissipation Factor (D)
Display hold
Relative mode
Tolerance mode
RS 232 Interface (cable and software required)
 
General Specification
Items:
Model 879
Model 878A
Parameters Measured
L/C/R/D/Q/Θ
L/C/R/D/Q
Measuring Circuit Mode
Inductance (L)
–Defaults to series mode
Capacitance/ Resistance (C/R)
-Defaults to parallel mode
Displays
L/C/R: Maximum display 19999
D/Q: Maximum display 999 (Auto Range).
Ranging Mode
Auto & Manual
Measuring Terminals
3 terminals with sockets
Test Frequency
Accuracy:
±0.1 %
100Hz=100 Hz
120Hz= 120 Hz
1KHz =1010 Hz
10KHz= 9.6 KHz
120Hz= 120 Hz
1KHz =1010 Hz
Backlit display
Included
Excluded
Tolerance mode
1%, 5%, 10%, 20%
1%, 5%, 10%
Test Signal Level
0.6Vrms approx.
Measuring Rate
1 time/second, nominal
Response time
Approx. 1 second/ DUT (device under test)(@ manual range)
Auto Power-Off
5 minutes approx. without operation
Temperature Coefficient
0.15 x (Specified Accuracy) / ℃(0-18℃ or 28-40℃)
Operation Temperature
0℃ to 40℃; 0-70% R.H.
Storage Temperature
-20℃ to +50℃; 0-80% R.H.
Low Battery Indication
Approx. 6.8V
Power Consumption
Approx. 40mA for operation/ 0.08mA after Auto Power-off.
Power Requirements
1) DC 9V Battery
2) Ext. DC Adaptor: DC 12Vmin –15Vmax. (Load 50mA Min.)
Protective Fuse
0.1A/250V Fuse (input protective)
Standard Accessories
Test alligator clips (pair)
DC 9V Battery.
User manual
Dimensions (L/W/H)
7.2/ 3.4/ 1.6" (184/ 87/ 41 mm)
Weight
11.6oz (330 grams)
 
 
 
Electrical Specification
Accuracy is expressed as: ± (% of reading + no. of least significant digits) at 23℃±5℃and <75% R.H.
Model 879:
o Resistance (Parallel mode)
Test Frequency: 100 / 120 Hz
Range
Maximum Display
Accuracy
Specified Note
@100 Hz
@120Hz
10MW
9.999MW
2.0%+8 *3
2.0%+8 *3
After open cal.
2000KW
1999.9KW
0.5%+5
0.5%+5
After open cal.
200KW
199.99KW
0.5%+3
0.5%+3
-
20KW
19.999KW
0.5%+3
0.5%+3
-
2000W
1999.9W
0.5%+3
0.5%+3
-
200W
199.99W
0.8%+5
0.8%+5
After short cal.
20W
19.999W
1.2%+40
1.2%+40
After short cal.
Test Frequency: 1K / 10K Hz
Range
Maximum Display
Accuracy
Specified Note
@1K Hz
@10KHz
10 MW
9.999MW
2.0%+8 * 3
3.5%+10 * 3
After open cal.
2000 KW
1999.9KW
0.5%+5
2.0%+10
After open cal.
200 KW
199.99KW
0.5%+3
1.5%+5
-
20 KW
19.999KW
0.5%+3
1.5%+5
-
2000 W
1999.9 W
0.5%+3
1.5%+5
-
200 W
199.99 W
0.8%+5
2.0%+10
After short cal.
20 W
19.999 W
1.2%+40
2.5%+200
After short cal.
Notes:
1.      This specification is based on the measurement performed at the test socket.
2.      DUT (Device Under Test) & Test leads be properly shielded to GUARD if necessary.
3.      This specification is based on battery operation.
o Capacitance (Parallel mode) Test Frequency: 100 / 120 Hz
Range
Maximum Display
Accuracy
Spec. Note
Capacitance
DF
10mF
19.99mF *5
3.0%+5
(DF<0.1)
10%+100/Cx+5 (DF<0.1)
After short cal.
1000μF
1999.9μF*6
1.0%+5
(DF<0.1)
2%+100/Cx+5 (DF<0.1)
After short cal.
200μF
199.99μF
0.7%+3
DF<0.5
0.7%+100/Cx+5 (DF<0.5)
-
20μF
19.999μF
0.7%+3
(DF<0.5)
0.7%+100/Cx+5 (DF<0.5)
-
2000nF
1999.9nF
0.7%+3
(DF<0.5)
0.7%+100/Cx+5 (DF<0.5)
-
200nF
199.99nF
0.7%+5
(DF<0.5)
0.7%+100/Cx+5 (DF<0.5)
After open cal.
20nF
19.999nF
1.0%+5
(DF<0.1)
2%+100/Cx+ (DF<0.1)
After open cal.
Test Frequency: 1 KHz
Range
Maximum Display
Accuracy
Spec. Note
Capacitance
DF
1mF
1.999mF *5
3.0%+5
(DF<0.1)
10%+100/Cx+5 (DF<0.1)
After short cal.
200μF
199.99μF
1.0%+5
(DF<0.1)
2.0%+100/Cx+5 (DF<0.1)
After short cal.
20μF
19.999μF
0.7%+3
(DF<0.5)
0.7%+100/Cx+5 (DF<0.5)
-
2000nF
1999.9nF
0.7%+3
(DF<0.5)
0.7%+100/Cx+5 (DF<0.5)
-
200nF
199.99nF
0.7%+3
(DF<0.5)
0.7%+100/Cx+5 (DF<0.5)
-
20nF
19.999nF
0.7%+5
(DF<0.5)
0.7%+100/Cx+5 (DF<0.5)
After open cal.
2000pF
1999.9pF
1.0%+5
(DF<0.1)
2.0%+100/Cx+5 (DF<0.1)
After open cal.
Test Frequency: 10 KHz
Range
Maximum Display
Accuracy
Spec. Note
Capacitance
DF
50μF
50.0μF
3.0%+8
(DF<0.1)
12%+100/Cx+10 (DF<0.1)
After short cal.
20μF
19.999μF
3.0%+6
(DF<0.2)
5.0%+100/Cx+8 (DF<0.2)
After short cal.
2000nF
1999.9nF
1.5%+5
(DF<0.5)
1.5%+100/Cx+6 (DF<0.5)
-
200nF
199.99nF
1.5%+5
(DF<0.5)
1.5%+100/Cx+6 (DF<0.5)
-
20nF
19.999nF
1.5%+5
(DF<0.5)
1.5%+100/Cx+6 (DF<0.5)
-
2000pF
1999.9pF
2.0%+6
(DF<0.5)
3.0%+100/Cx+6 (DF<0.5)
After open cal.
200pF
199.99pF
3.0%+8
(DF<0.1)
5.0%+100/Cx+8 (DF<0.1)
After open cal.
Notes:
1.    Q Value is the reciprocal of DF.
2.    This specification is based on the measurement performed at the test socket.
3.    DUT & Test leads should be properly shielded to GUARD if necessary.
4.    Cx=Counts of displayed C value, e.g. C=88.88μF then Cx=8888.
5.    This reading can be extended to 1999 MAX display with accuracy not specified.
6.    This reading can be extended to 19999 MAX display with accuracy not specified.
oInductance(Series mode)
Test Frequency: 100 / 120Hz
Range
Maximum
Display
Accuracy (DF<0.5)
Spec. Note
Inductance
DF
1000H
999.9H
1.0%+(Lx
/10000) %+5
2%+100/Lx+5
After open cal.
200H
199.99H
0.7%+(Lx
/10000)%+5
1.2%+100/Lx+5
-
20H
19.999H
0.7%+(Lx
/10000)%+5
1.2%+100/Lx+5
-
2000m
1999.9mH
0.7%+(Lx
/10000)%+5
1.2%+100/Lx+5
-
200mH
199.99mH
1.0%+(Lx
/10000)%+5
3%+100/Lx+5
After short cal.
20mH
19.999mH
2.0%+(Lx
/10000)%+5
10%+100/Lx+5
After short cal.
Test Frequency: 1 KHz
Range
Maximum Display
Accuracy (DF<0.5)
Spec. Note
Inductance
DF
100H
99.99H
1.0%+(Lx
/10000) %+5
2.0%+100/Lx+5
After open cal.
20H
19.999H
0.7%+(Lx
/10000)%+5
1.2%+100/Lx+5
-
2000mH
1999.9mH
0.7%+(Lx
/10000)%+5
1.2%+100/Lx+5
-
200mH
199.99mH
0.7%+(Lx
/10000)%+5
1.2%+100/Lx+5
-
20mH
19.999mH
1.0%+(Lx
/10000)%+5
3.0%+100/Lx+5
After short cal.
2000mH
1999.9mH
2.0%+
(Lx/10000)%+5
10%+100/Lx+5
After short cal.
Test Frequency: 10 KHz
Range
Maximum Display
Accuracy (DF<0.5)
Spec. Note
Inductance
DF
1000mH
999.9mH
2.0%+(Lx
/10000)%+ 8
2.0%+100/Lx+10
-
200mH
199.99mH
1.5%+(Lx
/10000)%+8
2.0%+100/Lx+10
-
20mH
19.999mH
1.5%+(Lx /10000)%+10
3.0 %+100/Lx+15
-
2000mH
1999.9mH
2.0%+(Lx /10000)%+10
8.0 %+100/Lx+20
After short cal.
Notes:
1.    1.Q Value is the reciprocal of DF.
2.    This specification is based on the measurement performed at the test socket.
3.    DUT & Test leads shall be properly shielded to GUARD if necessary.
4.    Lx=counts of displayed L value, e.g. L=88.88H, then Lx=8888.
 
Model 878A:
oResistance (parallel mode)
Test Frequency: 120Hz / 1KHz
Range
Maximum Display
Accuracy
Specified Note
@120 Hz
@1KHz
10MW
9.999MW
2.0%+8 *3
2.0%+8 *3
After open cal.
2000KW
1999.9KW
0.5%+5
0.5%+5
After open cal.
200KW
199.99KW
0.5%+3
0.5%+3
-
20KW
19.999KW
0.5%+3
0.5%+3
-
2000W
1999.9W
0.5%+3
0.5%+3
-
200W
199.99W
0.8%+5
0.8%+5
After short cal.
20W
19.999W
1.2%+40
1.2%+40
After short cal.
Notes:
1.      This specification is based on the measurement performed at the test socket.
2.      DUT (Device Under Test) & Test leads be properly shielded to GUARD if necessary.
3.      This specification is based on battery operation.
o Capacitance (parallel mode) Test Frequency: 120 Hz
Range
Maximum Display
Accuracy
Spec. Note
Capacitance
DF
10mF
19.99mF *5
3.0%+5
(DF<0.1)
10%+100/Cx+5 (DF<0.1)
After short cal.
1000μF
1999.9μF*6
1.0%+5
(DF<0.1)
2%+100/Cx+5 (DF<0.1)
After short cal.
200μF
199.99μF
0.7%+3
DF<0.5
0.7%+100/Cx+5 (DF<0.5)
-
20μF
19.999μF
0.7%+3
(DF<0.5)
0.7%+100/Cx+5 (DF<0.5)
-
2000nF
1999.9nF
0.7%+3
(DF<0.5)
0.7%+100/Cx+5 (DF<0.5)
-
200nF
199.99nF
0.7%+5
(DF<0.5)
0.7%+100/Cx+5 (DF<0.5)
After open cal.
20nF
19.999nF
1.0%+5
(DF<0.1)
2%+100/Cx+ (DF<0.1)
After open cal.
Test Frequency: 1 KHz
Range
Maximum Display
Accuracy
Spec. Note
Capacitance
DF
1mF
1.999mF *5
3.0%+5
(DF<0.1)
10%+100/Cx+5 (DF<0.1)
After short cal.
200μF
199.99μF
1.0%+5
(DF<0.1)
2.0%+100/Cx+5 (DF<0.1)
After short cal.
20μF
19.999μF
0.7%+3
(DF<0.5)
0.7%+100/Cx+5 (DF<0.5)
-
2000nF
1999.9nF
0.7%+3
(DF<0.5)
0.7%+100/Cx+5 (DF<0.5)
-
200nF
199.99nF
0.7%+3
(DF<0.5)
0.7%+100/Cx+5 (DF<0.5)
-
20nF
19.999nF
0.7%+5
(DF<0.5)
0.7%+100/Cx+5 (DF<0.5)
After open cal.
2000pF
1999.9pF
1.0%+5
(DF<0.1)
2.0%+100/Cx+5 (DF<0.1)
After open cal.
Notes:
1.    Q Value is the reciprocal of DF.
2.    This specification is based on the measurement performed at the test socket.
3.    DUT & Test leads should be properly shielded to GUARD if necessary.
4.    Cx=Counts of displayed C value, e.g. C=88.88μF then Cx=8888.
5.    This reading can be extended to 1999 MAX display with accuracy not specified.
6.    This reading can be extended to 19999 MAX display with accuracy not specified.
 
oInductance (Series mode)
Test Frequency: 120Hz
Range
Maximum
Display
Accuracy (DF<0.5)
Spec. Note
Inductance
DF
1000H
999.9H
1.0%+(Lx
/10000) %+5
2%+100/Lx+5
After open cal.
200H
199.99H
0.7%+(Lx
/10000)%+5
1.2%+100/Lx+5
-
20H
19.999H
0.7%+(Lx
/10000)%+5
1.2%+100/Lx+5
-
2000m
1999.9mH
0.7%+(Lx
/10000)%+5
1.2%+100/Lx+5
-
200mH
199.99mH
1.0%+(Lx
/10000)%+5
3%+100/Lx+5
After short cal.
20mH
19.999mH
2.0%+(Lx
/10000)%+5
10%+100/Lx+5
After short cal.
Test Frequency: 1 KHz
Range
Maximum Display
Accuracy (DF<0.5)
Spec. Note
Inductance
DF
100H
99.99H
1.0%+(Lx
/10000) %+5
2.0%+100/Lx+5
After open cal.
20H
19.999H
0.7%+(Lx
/10000)%+5
1.2%+100/Lx+5
-
2000mH
1999.9mH
0.7%+(Lx
/10000)%+5
1.2%+100/Lx+5
-
200mH
199.99mH
0.7%+(Lx
/10000)%+5
1.2%+100/Lx+5
-
20mH
19.999mH
1.0%+(Lx
/10000)%+5
3.0%+100/Lx+5
After short cal.
2000mH
1999.9mH
2.0%+
(Lx/10000)%+5
10%+100/Lx+5
After short cal.
 

产品编号 产品名称 产品简介 产品价格
ME-01879-00 BK878A通用LCR测试器 The 878A measures capacitance, resistance (of non-inductive components) and inductance. Components can be measured with selectable test frequencies in series or parallel modes as desired. 0¥

相关产品