  • 产品名称:KSD-400SKSD-400S 400升灭菌烘干箱
  • 定货号:KJ-00851-00
  • 推荐:
Product Name :
KSD-400S 
Product Description :
400 Litre Steriliser Dry Heat, Fan Assist Circulation. New 2009 
Nominal Tank Capacity :
400 litre chamber Working capacity of the chamber can be smaller.
Size selector :
«Smaller  Larger»400
Product Specification :
This Clifton Steriliser features precision time and temperature control with programming options suitable for hot-air sterilisation of containers and instruments up to 250°C.
Sterilser features door locking and chamber vent closure when starting and during a sterilisation program.
Microprocessor control with auto diagnostics, digital LCD display to 0.1°C resolution, twenty program memory with event log of actual temperature and time data with RS232 interface.
Featuring wipe clean touch pad controls illuminated LCD display for viewing wide range of settings, programs, temperature and time values.
This model a 400 litre capacity is supplied with 3 wire shelves, with the option for additional wire or perforated shelves up to maximum capacity of 14.
The high grade 304 spec stainless steel chamber is sealed shut with a solid door. Alternative door configuration includes a window in the door allowing chamber observation.
Temperature
Range: ambient +5°C - 250°C.
Microprocessor control, dual pt100 temperature probes11.
Auto diagnostic system.
Digital LCD display to 0.1°C resolution.
Sensitivity @ 70°C ±0.5°C On wire shelf in middle chamber.
Uniformity @ 150°C ±1.6°C Measured in a vertical axis in the chamber using Wired Shelves.
Calibration Certificate @ 170°C.
Temperature and time programmable
Temperature and time programmable in two ways:
Six segment temperature-time profile with delay start13, heating up9, hold at set point12, each with timers from 1 minute to 99 hours.
Continuous mode at set point with either delay start 13, or timed up to 999 hours12.
Date and time setting, program can start unsupervised13.
Audible alarm and display, setting 0.5°C to 10°C about set point10.
Event log and program review temperature and time records.
Average minimum and maximum temperature values recorded for each segment of the program for reference and validation.
Forced air circulation with Fan speed regulation from 10 to 100%.
RS232 interface.
Over-temperature protection
Cass 1, thermal cut out 260°C 7 and separate element overheat protection.
Class 2, experiment/sample protection, adjustable setting above set point, heating switched off, audible and visual indication8.
Accessory options Class 3, temperature protection, self resetting cut outs.
Temperature sensor fail alarm11.
Power failure control with auto resume.
Interior and Case
Using a 304 spec stainless steel interior for long working lfe and easy cleaning.
Case construction stainless steel with powder painted exterior.
Sterilisers up to 112 litres are stackable.
 
Internal Dimensions :
800w x 510d x 1040h mm
External Dimensions :
1015w x 710d x 1385h mm
Temperature Range :
ambient +5°C to 250°C
Sensitivity/temperature stability @45°C :
±0.5°C @ 70°C / ±0.7°C @ 150°C Sensitivity measurements based on temperature readings on shelf in centre of the chamber.
Uniformity @45°C :
±1.6°C @ 150°C Uniformity measured in vertical axis in the chamber using Wired Shelves.
Temperature setting/control :
Wipe clean controls/Keypad setting/Digital LCD display/Microprocessor control.
Temperature display resolution :
Fast realtime display of actual measured temperature in increments of ±0.1°C.
Warranty Period :
3 Years
Safety features :
Over-temperature protection
Cass 1, thermal cut out 260°C 7 and separate element overheat protection.
Class 2, experiment/sample protection, adjustable setting above set point, heating switched off, audible and visual indication8.
Accessory options Class 3, temperature protection, self resetting cut outs.
Temperature sensor fail alarm11.
Power failure control with auto resume.
Safety Class :
1 + 2 fitted, options for 3+
Power consumption (W) @ 220-240V:
3900W
Voltage :
230V 50-60Hz
Packed Weight :
250kg
Packed Dimensions :
120x88x180cm
Notes :
  A range of ‘factory fit’ options are listed below as Accessories.
  Before ordering please check the room or building access where this unit will be installed and electrical power requirements.
  Sensitivity measurements based on temperature readings on shelf in centre of the chamber.
  Uniformity measured in vertical axis in the chamber using Wired Shelves.
  Working capacity of chamber can be smaller.
  With the solid door design a viewing window can be specified.
  Sterilisers up to 112 litres are stackable.
  Do not place in direct sunlight and allow a 100mm air gap away from wall at rear. Ensure there is a minimum 300mm air gap from top of the unit to ceiling.
  When no programs are active, the display shows current time and temperature inside chamber. When running a program the display shows program number, temperature inside chamber and status symbol showing the status of heating.
 