Selectable 120 000/40 000/4 000-count multimeter 
Basic accuracy 0.012% V DC 
Voltage masurement up to 1000V DC et 750V AC 
Current measurement up to 12A 
True RMS AC, AC+DC with 100KHz bandwidth 
Resistance measurement up to 120Mohms 
Frequency measurement from 5Hz to 1MHz 
Selectable measurement rates 
2 or 4 wires resistance measurements 
dBm, %, MIN, MAX, HOLD, COMP functions 
Dual display 
RS-232 interface 
 
GPIB version available (BK5482GPIB)
Specifications assumptions:
One-year calibration cycle. Operating temperature at 18 °C to 28°C (64.4°F to 82.4°F). Accuracy is expressed as: ± % of reading + digits) after 30 minutes warm-up. Temperature coefficient: Add ± [0. 15 x (the applicable accuracy)/°C] for 0°C to 18°C and 28°C to 50°C. Relative Humidity (RH) up to 80% for 0 °C to 28°C (75% RH for 12M  Ohm and above ranges of resistance measurement); up to 70% for 28°C to 35°C; up to 50% for 35°C to 50°C All specifications are specified under single display mode in operation only. 
Display Counts and Reading Rates
Full Scale Display Counts
| 
 Slow  | 
 Medium  | 
 Fast  | 
| 
 119,999  | 
 39,999  | 
 3,999  | 
Reading Rates on Single Display (Readings/Sec)
| 
 Measurement Functions  | 
 Slow  | 
 Medium  | 
 Fast  | 
| 
 DCV  | 
 2  | 
 5  | 
 20  | 
| 
 DCA  | 
 2  | 
 5  | 
 20  | 
| 
 Diode  | 
 2  | 
 5  | 
 20  | 
| 
 ACV  | 
 2  | 
 4.2  | 
 20  | 
| 
 ACA  | 
 2  | 
 4.2  | 
 20  | 
| 
 2-wires  Ohm  | 
 2  | 
 4  | 
 17  | 
| 
 4-wires  Ohm 4M Ohm/1.2M Ohm range and below  | 
 0.6  | 
 0.8  | 
 0.9  | 
| 
 4-wires  Ohm 12M Ohm range and above  | 
 1.4  | 
 1.7  | 
 1.9  | 
| 
 Frequency  | 
 1.2  | 
 1.7  | 
 2.4  | 
| 
 ACV+DCV  | 
 0.4  | 
 0.5  | 
 0.7  | 
| 
 ACA+DCA  | 
 0.4  | 
 0.5  | 
 0.7  | 
Reading Rates on Dual Display (Readings/Sec)
| 
 Measurement Functions  | 
 Slow  | 
 Medium  | 
 Fast  | 
| 
 DCV / ACV  | 
 0.7  | 
 1.0  | 
 1.9  | 
| 
 DCA / ACA  | 
 0.7  | 
 1.0  | 
 1.9  | 
| 
 DCV / DCA  | 
 0.7  | 
 1.0  | 
 1.9  | 
| 
 DCV / ACA  | 
 0.2  | 
 0.2  | 
 0.5  | 
| 
 ACV / ACA  | 
 0.2  | 
 0.2  | 
 0.5  | 
| 
 ACV / DCA  | 
 0.7  | 
 1.0  | 
 1.9  | 
| 
 ACV / Frequency  | 
 0.5  | 
 0.7  | 
 1.1  | 
| 
 ACA / Frequency  | 
 0.6  | 
 0.8  | 
 1.3  | 
| 
 ACV+DCV / DCV  | 
 0.5  | 
 0.6  | 
 0.9  | 
| 
 ACA+DCA / DCV  | 
 0.1  | 
 0.2  | 
 0.4  | 
| 
 ACA+DCA / ACV  | 
 0.1  | 
 0.2  | 
 0.4  | 
| 
 ACA+DCA / DCA  | 
 0.5  | 
 0.6  | 
 0.9  | 
| 
 dBm(ACV) / Reference  Ohm  | 
 2.1  | 
 4.2  | 
 11.9  | 
| 
 dBm(ACV) / ACV  | 
 0.7  | 
 1.0  | 
 1.7  | 
| 
 dBm(ACV) / DCV  | 
 0.6  | 
 1.0  | 
 1.7  | 
| 
 dBm(ACV) / Frequency  | 
 0.7  | 
 1.0  | 
 1.7  | 
Note 1: Above table shows some common combinations and applications of using dual display.
Note 2: Using RS-232 or GPIB remote interface, the reading rate approximates to normal mode.
DC Voltage
Resolution, Full Scale Reading and Accuracy
| 
 Rate (1)  | 
 Range  | 
 Resolution  | 
 Full Scale Reading  | 
 Accuracy (1 year)  | 
 Typical Input Impedance (4)  | 
| 
 S  | 
 120mV 1.2V 12V 120V 1000V  | 
 1uV 10uV 100uV 1mV 10mV  | 
 119.999 1.19999 11.9999 119.999 1000.00(3)  | 
 0.012% + 2 0.012% + 2 0.012% + 2 0.012% + 2 0.012% + 2  | 
 10.0M Ohm 10.0M Ohm 11.1M Ohm 10.1M Ohm 10.0M Ohm  | 
| 
 M  | 
 400mV 4V 40V 400V 1000V  | 
 10uV 100uV 1mV 10mV 100mV  | 
 399.99 3.9999 39.999 399.99 1000.0(3)  | 
 0.012% + 5 0.012% + 5 0.012% + 5 0.012% + 5 0.012% + 5  | 
 10.0M Ohm 11.1M Ohm 10.1M Ohm 10.0M Ohm 10.0M Ohm  | 
| 
 F  | 
 400mV 4V 40V 400V 1000V  | 
 100uV 1mV 10mV 100mV 1V  | 
 399.9 3.999 39.99 399.9 1000(3)  | 
 0.012% + 8 (2) 0.012% + 5 0.012% + 5 0.012% + 5 0.012% + 5  | 
 10.0M Ohm 11.1M Ohm 10.1M Ohm 10.0M Ohm 10.0M Ohm  | 
| 
 (1) Rate: S (Slow), M (Medium), and F (Fast). (2) Use relative (REL) modifier. (3) In Vdc 1000V range, 5% over-range (1050Vdc) is readable. (4) Input Impedance is in paralleled with capacitance <120pF.  | 
Maximum input voltage: 1000Vdc or peak ac on any range Response Time: Approximately 1.0 second when the displayed reading reaches 99.9% dc value of the tested input signal at the same range. 
Note: When voltage (ac+dc) measurement function is selected, the Vdc input impedance is paralleled with an ac-coupled 1.1M Ohm ac divider.
Noise Rejection Ratio
| 
 Rate  | 
 CMRR (1)  | 
 NMRR (2)  | 
| 
 S / M / F  | 
 >90dB at dc, 50/60Hz ± 0.1% (1k OhmUnbalanced)  | 
 >50dB at 50/60Hz ± 0.1%  | 
| 
 (1) CMRR is the Common Mode Reject Ratio (2) NMRR is the Normal Mode Rejection Ratio  | 
AC Voltage (True RMS, AC Coupling Mode)
Resolution and Full Scale Reading
| 
 Range  | 
 Resolution  | 
 Full Scale Reading  | 
| 
 S  | 
 M & F  | 
 S  | 
 M  | 
 F  | 
 S  | 
 M  | 
 F  | 
| 
 120mV 1.2V 12V 120V 750V  | 
 400mV 4V 40V 400V 750V  | 
 1uV 10uV 100uV 1mV 10mV  | 
 10uV 100uV 1mV 10mV 100mV  | 
 100uV 1mV 10mV 100mV 1V  | 
 119.999 1.19999 11.9999 119.999 750.00 (1)  | 
 399.99 3.9999 39.999 399.99 750.0 (1)  | 
 399.9 3.999 39.99 399.9 750 (1)  | 
| 
 (1) In Vac 750V range, 5% over-range (787.5V rms) is readable  | 
Accuracy
| 
 Rate  | 
 Range  | 
 Accuracy (1 year) (1)  | 
| 
 20 to 45 Hz  | 
 45 to 10k Hz  | 
 10 to 30 kHz  | 
 30 to 100 kHz  | 
| 
 S  | 
 120.000mV 1.20000V 12.0000V 120.000V 750.00V  | 
 1% + 5 1% + 5 1% + 5 1% + 5 (2) 1% + 5 (2)  | 
 0.2% + 5 0.2% + 5 0.2% + 5 0.2% + 5 0.2% + 5  | 
 1.5% + 10 1% + 5 1% + 5 1% + 5 1% + 5  | 
 5% + 15 3% + 10 3% + 10 3% + 10 3% + 10 (3)  | 
| 
 M  | 
 400.00mV 4.0000V 40.000V 400.00V 750.0V  | 
 1% + 40 1% + 40 1% + 40 1% + 40 (2) 1% + 40 (2)  | 
 0.2% + 40 0.2% + 40 0.2% + 40 0.2% + 40 0.2% + 40  | 
 1.5% + 80 1% + 40 1% + 40 1% + 40 1% + 40  | 
 5% + 120 3% + 80 3% + 80 3% + 80 3% + 80 (3)  | 
| 
 F  | 
 400.0mV 4.000V 40.00V 400.0V 750V  | 
 1% + 100 1% + 100 1% + 100 1% + 100 1% + 100 (2)  | 
 0.2% + 100 0.2% + 100 0.2% + 100 0.2% + 100 0.2% + 100  | 
 1.5% + 300 1% + 100 1% + 100 1% + 100 1% + 100  | 
 5% + 300 3% + 200 3% + 200 3% + 200 3% + 200 (3)  | 
| 
 (1) Specified accuracy at input >5% of full scale  | 
| 
 (2) For input <200V rms  | 
 (3) For input <500V rms  | 
| 
    | 
    | 
    | 
    | 
    | 
    | 
    | 
Measurement method: True RMS Maximum Crest Factor: 3.0 at full scale Maximum input voltage: 750V rms, 1100V peak ac
2x107 V-Hz product on any range, normal mode input
1x106 V-Hz product on any range, common mode input Input Impedance: 1M Ohm in parallel with capacitance <120pF Response Time: Approximately 1.5 seconds when the displayed reading reaches 99.9% ac rms value of the tested input signal at the same range. 
AC Voltage (True RMS, AC+DC Coupling Mode)
Resolution and Full Scale Reading
| 
 Range (1)  | 
 Resolution  | 
 Full Scale Reading  | 
| 
 S  | 
 M & F  | 
 S  | 
 M  | 
 F  | 
 S  | 
 M  | 
 F  | 
| 
 120mV 1.2V 12V 120V 750V  | 
 400mV 4V 40V 400V 750V  | 
 1uV 10uV 100uV 1mV 10mV  | 
 10uV 100uV 1mV 10mV 100mV  | 
 100uV 1mV 10mV 100mV 1V  | 
 119.999 1.19999 11.9999 119.999 750.00 (2)  | 
 399.99 3.9999 39.999 399.99 750.0 (2)  | 
 399.9 3.999 39.99 399.9 750 (2)  | 
| 
 (1) Vdc and Vac are automatically set at the same range (2) In Vac 750V range, 5% over-range (787.5V rms) is readable  | 
Accuracy
| 
 Rate  | 
 Range  | 
 Accuracy (1 year) (1)  | 
| 
 45 to 10k Hz  | 
 10 to 30 kHz  | 
 30 to 100 kHz  | 
| 
 S  | 
 120.000mV 1.20000V 12.0000V 120.000V 750.00V  | 
 0.2% + 7 0.2% + 7 0.2% + 7 0.2% + 7 0.2% + 7  | 
 1.5% + 12 1% + 7 1% + 7 1% + 7 1% + 7  | 
 5% + 18 3% + 12 3% + 12 3% + 12 3% + 12 (2)  | 
| 
 M  | 
 400.00mV 4.0000V 40.000V 400.00V 750.0V  | 
 0.2% + 45 0.2% + 43 0.2% + 43 0.2% + 43 0.2% + 43  | 
 1.5% + 83 1% + 43 1% + 43 1% + 43 1% + 43  | 
 5% + 125 3% + 83 3% + 83 3% + 83 3% + 83 (2)  | 
| 
 F  | 
 400.0mV 4.000V 40.00V 400.0V 750V  | 
 0.2% + 100 0.2% + 100 0.2% + 100 0.2% + 100 0.2% + 100  | 
 1.5% + 300 1% + 100 1% + 100 1% + 100 1% + 100  | 
 5% + 300 3% + 200 3% + 200 3% + 200 3% + 200 (2)  | 
| 
 (1) Specified accuracy at input >5% of full scale  | 
 (2) For input <500V rms  | 
| 
    | 
    | 
    | 
    | 
    | 
    | 
Measurement method: True RMS AC+DC Maximum Crest Factor: 3.0 at full scale Maximum input voltage: 750V rms, 1100V peak ac
2x107 V-Hz product on any range, normal mode input
1x106 V-Hz product on any range, common mode input Input Impedance: 1M Ohm in parallel with capacitance <120pF Response Time: Approximately 2.5 seconds when the displayed reading reaches 99.9% (ac+dc) rms value of the tested input signal at the same range. 
DC Current
| 
 Rate  | 
 Range  | 
 Resolution  | 
 Full Scale Reading  | 
 Accuracy (1 year)  | 
 Burden Voltage (1) & Shunt Resistor  | 
| 
 S  | 
 12mA 120mA 1200mA 12A  | 
 0.1uA 1mA 10mA 100mA  | 
 11.9999 119.999 1199.99 11.9999  | 
 0.1% + 2 0.1% + 2 0.15% + 2 0.2% + 2  | 
 <0.15V / 10 Ohm <1.5V / 10 Ohm <0.3V / 0.1 Ohm <0.6V / 0.01 Ohm  | 
| 
 M  | 
 40mA 120mA 1200mA 12A  | 
 1uA 10uA 100uA 1mA  | 
 39.999 119.99 1199.9 11.999  | 
 0.1% + 6 0.1% + 3 0.15% + 3 0.2% + 3  | 
 <0.5V / 10 Ohm <1.5V / 10 Ohm <0.3V / 0.1 Ohm <0.6V / 0.01 Ohm  | 
| 
 F  | 
 40mA 120mA 1200mA 12A  | 
 10uA 100uA 1mA 10mA  | 
 39.99 119.9 1199 11.99  | 
 0.05% + 15 (2) 0.05% + 5 0.15% + 5 0.2% + 5  | 
 <0.5V / 10 Ohm <1.5V / 10 Ohm <0.3V / 0.1 Ohm <0.6V / 0.01 Ohm  | 
| 
 (1) Typical at full scale reading and voltage across the input terminals (2) Use relative (REL) modifier  | 
Maximum Input and Overload Current Protection (for dc and ac current):
mA Input Terminal: 1200mA dc or ac rms and protected with 2A/250V, IEC-127 sheet 1 fast blow fuse;
12A input terminal: 10A dc or ac rms continuous, or 12A dc or ac rms for 30 seconds maximum and protected with 15A/500V, breaking capacity 10,000A fast blow fuse. Response Time: Approximately 1.0 second when the displayed reading reaches 99.9% dc value of the tested input signal at the same range. 
AC Current (True RMS, AC Coupling Mode)
Resolution, Full Scale Reading and Burden Voltage
| 
 Rate  | 
 Range  | 
 Resolution  | 
 Full Scale Reading  | 
 Burden Voltage (1) & Shunt Resistor  | 
| 
 S  | 
 12mA 120mA 1200mA 12A  | 
 0.1uA 1uA 10uA 100uA  | 
 11.9999 119.999 1199.99 11.9999  | 
 <0.15V / 10 Ohm <1.5V / 10 Ohm <0.3V / 0.1 Ohm <0.6V / 0.01 Ohm  | 
| 
 M  | 
 40mA 120mA 1200mA 12A  | 
 1uA 10uA 100uA 1mA  | 
 39.999 119.99 1199.9 11.999  | 
 <0.5V / 10 Ohm <1.5V / 10 Ohm <0.3V / 0.1 Ohm <0.6V / 0.01 Ohm  | 
| 
 F  | 
 40mA 120mA 1200mA 12A  | 
 10uA 100uA 1mA 10mA  | 
 39.99 119.9 1199 11.99  | 
 <0.5V / 10 Ohm <1.5V / 10 Ohm <0.3V / 0.1 Ohm <0.6V / 0.01 Ohm  | 
| 
 (1) Typical at full scale reading and rms voltage across the input terminals  | 
Accuracy
| 
 Rate  | 
 Range  | 
 Accuracy (1 year) (1)  | 
| 
 20 to 45 Hz  | 
 45 to 2k Hz  | 
 2 to 10 kHz  | 
| 
 S  | 
 12mA 120mA 1200mA 12A  | 
 1.5% + 5 1.5% + 2 1.5% + 2 2% + 2 (<1.2A)  | 
 0.5% + 5 0.5% + 2 0.5% + 2 1% + 2  | 
 2% + 10 2% + 5 2% + 5 -  | 
| 
 M  | 
 40mA 120mA 1200mA 12A  | 
 1.5% + 40 1.5% + 12 1.5% + 12 2% + 12 (<1.2A)  | 
 0.5% + 40 0.5% + 12 0.5% + 12 1% + 12  | 
 2% + 80 2% + 30 2% + 30 -  | 
| 
 F  | 
 40mA 120mA 1200mA 12A  | 
 1.5% + 100 1.5% + 100 1.5% + 100 2% + 100 (<1.2A)  | 
 0.5% + 100 0.5% + 100 0.5% + 100 1% + 100  | 
 2% + 200 2% + 200 2% + 200 -  | 
| 
 (1) Specified accuracy at input >5% of full scale  | 
Measurement method: True RMS Maximum Crest Factor: 3.0 at full scale Response Time: Approximately 1.5 seconds when the displayed reading reaches 99.9% ac rms value of the tested input signal at the same range. 
AC Current (True RMS, AC+DC Coupling Mode)
Resolution, Full Scale Reading and Burden Voltage
| 
 Rate  | 
 Range  | 
 Resolution  | 
 Full Scale Reading  | 
 Burden Voltage (1) & Shunt Resistor  | 
| 
 S  | 
 12mA 120mA 1200mA 12A  | 
 0.1uA 1uA 10uA 100uA  | 
 11.9999 119.999 1199.99 11.9999  | 
 <0.15V / 10 Ohm <1.5V / 10 Ohm <0.3V / 0.1 Ohm <0.6V / 0.01 Ohm  | 
| 
 M  | 
 40mA 120mA 1200mA 12A  | 
 1uA 10uA 100uA 1mA  | 
 39.999 119.99 1199.9 11.999  | 
 <0.5V / 10 Ohm <1.5V / 10 Ohm <0.3V / 0.1 Ohm <0.6V / 0.01 Ohm  | 
| 
 F  | 
 40mA 120mA 1200mA 12A  | 
 10uA 100uA 1mA 10mA  | 
 39.99 119.9 1199 11.99  | 
 <0.5V / 10 Ohm <1.5V / 10 Ohm <0.3V / 0.1 Ohm <0.6V / 0.01 Ohm  | 
| 
 (1) Typical at full scale reading and rms voltage across the input terminals  | 
Accuracy
| 
 Rate  | 
 Range  | 
 Accuracy (1 year) (1)  | 
| 
 45 to 2k Hz  | 
 2 to 10 kHz  | 
| 
 S  | 
 12mA 120mA 1200mA 12A  | 
 0.5% + 7 0.5% + 4 0.5% + 4 1% + 4  | 
 2% + 12 2% + 7 2% + 7 -  | 
| 
 M  | 
 40mA 120mA 1200mA 12A  | 
 0.5% + 42 0.5% + 15 0.5% + 15 1% + 15  | 
 2% + 80 2% + 30 2% + 30 -  | 
| 
 F  | 
 40mA 120mA 1200mA 12A  | 
 0.5% + 100 0.5% + 100 0.5% + 100 1% + 100  | 
 2% + 200 2% + 200 2% + 200 -  | 
| 
 (1) Specified accuracy at input >5% of full scale  | 
Measurement range: Vdc and Vac are automatically set at the same range Measurement method: True RMS AC+DC Maximum Crest Factor: 3.0 at full scale Response Time: Approximately 2.5 seconds when the displayed reading reaches 99.9% (ac+dc) rms value of the tested input signal at the same range. 
Resistance (2-wire  Ohm and 4-wire  Ohm)
| 
 Rate  | 
 Range (1)  | 
 Resolution  | 
 Full Scale Reading  | 
 Test Current  | 
 Accuracy (1 year)  | 
| 
 2-wire  | 
 4-wire  | 
| 
 S  | 
 120 Ohm 1.2k Ohm 12k Ohm 120k Ohm 1.2M Ohm 12M Ohm 120M Ohm  | 
 1m Ohm 10m Ohm 100m Ohm 1 Ohm 10 Ohm 100 Ohm 1k Ohm  | 
 119.999 1.19999 11.9999 119.999 1.19999 11.9999 119.999  | 
 0.5mA 0.5mA 100mA 10mA 1mA 100nA 10nA  | 
 0.1% + 2 (2) 0.08% + 2 0.06% + 2 0.06% + 2 0.15% + 2 1.5% + 2 5.0% + 2  | 
 0.05% + 2 0.05% + 2 0.05% + 2 0.05% + 2 0.15% + 2 1.5% + 2 5.0% + 2  | 
| 
 M  | 
 400 Ohm 4k Ohm 40k Ohm 400k Ohm 4M Ohm 40M Ohm 300M Ohm  | 
 10m Ohm 100m Ohm 1 Ohm 10 Ohm 100 Ohm 1K Ohm 10k Ohm  | 
 399.99 3.9999 39.999 399.99 3.9999 39.999 299.99  | 
 0.5mA 100mA 50mA 5mA 500nA 50nA 10nA  | 
 0.1% + 5 (2) 0.08% + 3 (2) 0.06% + 3 0.06% + 3 0.15% + 3 1.5% + 3 5.0% + 5  | 
 0.05% + 5 (2) 0.05% + 3 0.05% + 3 0.05% + 3 0.15% + 3 1.5% + 3 5.0% + 5  | 
| 
 F  | 
 400 Ohm 4k Ohm 40k Ohm 400k Ohm 4M Ohm 40M Ohm 300M Ohm  | 
 100m Ohm 1 Ohm 10 Ohm 100 Ohm 1k Ohm 10k Ohm 100k Ohm  | 
 399.9 3.999 39.99 399.9 3.999 39.99 299.9  | 
 0.5mA 100mA 50mA 5mA 500nA 50nA 10nA  | 
 0.1% + 8 (2) 0.08% + 5 (2) 0.06% + 5 (2) 0.06% + 5 0.06% + 5 0.3% + 5 3.0% + 8  | 
 0.05% + 8 (2) 0.05% + 5 (2) 0.05% + 5 0.05% + 5 0.05% + 5 0.3% + 5 3.0% + 8  | 
| 
 (1) In order to eliminate the noise interference, which might be induced to the test leads, it is recommended to use a shielded test cable for measuring resistance above 120K Ohm. (2) Use relative (REL) modifier.  | 
Open Circuit Voltage: < +5.0V dc Zeroing error: 0.05 Ohm or less (excluding test lead resistances) in each range when REL modifier is used Response time: Approximately 1.5 seconds for 12M Ohm and ranges below 12M Ohm; approximately 5 seconds for 40M Ohm range; approximately 10 seconds for 120M Ohm ; approximately 25 seconds for 300M Ohm range. Maximum Input Protection: 500V dc or ac rms 
Diode Test/Continuity
| 
 Rate  | 
 Maximum Reading  | 
 Resolution  | 
| 
 S  | 
 1.19999V  | 
 10mV  | 
| 
 M  | 
 2.4999V  | 
 100mV  | 
| 
 F  | 
 2.499V  | 
 1mV  | 
Open Circuit Voltage: < +5.0V dc Test Current: Approximately 0.5mA dc Audible Tone: Continuous beep for continuity and single tone for normal forward-biased diode or semiconductor junction Continuity level: Approximately below +50mVdc Maximum Input Protection: 500V dc or ac rms 
Resistance/Continuity (2-wire)
| 
 Rate  | 
 Range  | 
 Resolution  | 
 Maximum Reading  | 
 Accuracy  | 
| 
 S  | 
 120 Ohm  | 
 1m Ohm  | 
 119.999  | 
 0.1% + 2 (1)  | 
| 
 M  | 
 400 Ohm  | 
 10m Ohm  | 
 399.99  | 
 0.1% + 5 (1)  | 
| 
 F  | 
 400 Ohm  | 
 100m Ohm  | 
 399.9  | 
 0.1% + 8 (1)  | 
| 
 (1) Use relative (REL) modifier  | 
Open Circuit Voltage: < +5.0V dc Test Current: Approximately 0.5mA dc Audible Tone: Continuous beep for reading is less than 10 Ohm Zeroing error: 0.05 Ohm or less (excluding test lead resistances) in each range when REL modifier is used Maximum Input Protection: 500V dc or ac rms 
Frequency
Resolution, Full Scale Reading and Accuracy
| 
 Range (Hz)  | 
 Measurement Range (Hz)  | 
 Resolution (Hz)  | 
 Full Scale Reading  | 
 Accuracy (1 year) (1)  | 
 Input Sensitivity (Sine wave)  | 
| 
 1200  | 
 5 to 1200  | 
 10m  | 
 1199.99  | 
 0.005 + 3  | 
 40mV rms  | 
| 
 12k  | 
 10 to 12k  | 
 100m  | 
 11.9999  | 
 0.005 + 2  | 
| 
 120k  | 
 100 to 120k  | 
 1  | 
 119.999  | 
 0.005 + 2  | 
| 
 1M (2)  | 
 1k to 1M  | 
 10  | 
 1.19999  | 
 0.005 + 2  | 
 0.5V rms  | 
| 
 (1) Specified accuracy at input >5% of full scale (2) If tested frequency is greater than 1MHz, it will be displayed but no specified accuracy is guaranteed.  | 
Measurement method: True RMS Maximum Crest Factor: 3.0 at full scale Maximum input voltage: 750V rms, 1100V peak ac
2x107 V-Hz product on any range, normal mode input
1x106 V-Hz product on any range, common mode input Input Impedance: 1M Ohm in parallel with capacitance <120pF Response Time: Approximately 1.2 seconds when the displayed reading reaches 99.9% frequency value of the tested input signal at the same range. 
dBm (decibel calculation)
Reference Impedance (1)
| 
 2 Ohm (2) 4 Ohm (2) 8 Ohm (2) 16 Ohm (2)
     | 
 50 Ohm 75 Ohm 93 Ohm 110 Ohm 124 Ohm 125 Ohm  | 
 135 Ohm 150 Ohm 250 Ohm 300 Ohm 500 Ohm 600 Ohm (3)  | 
 800 Ohm 900 Ohm 1000 Ohm 1200 Ohm 8000 Ohm    | 
| 
 (1) Reference impedance is displayed on the secondary display (2) Reading displayed in watts (Audio Power) (3) Default reference impedance  | 
Range and Accuracy
| 
 Rate  | 
 Voltage Range (1,2)  | 
 Input Voltage  | 
 dBm (3) Range @ 600 Ohm Ref  | 
 Accuracy (dB)  | 
| 
 20 to 45 Hz  | 
 45 to 10 kHz  | 
 0 to 100 kHz  | 
| 
 S  | 
 120mV 1.2V 12V 120V  | 
 6mV ~ 120mV 120mV ~ 1.2V 1.2V ~ 12V 12V ~ 120V  | 
 -42.20 ~ -16.20 -16.20 ~ 3.80 3.80 ~ 23.80 23.80 ~ 43.80  | 
 1.0 0.8 0.8 0.8  | 
 0.2 0.1 0.1 0.1  | 
 1.0 0.8 0.8 0.8  | 
| 
 1000V (dc) 750V (ac)  | 
 120V ~ 1000V 120V ~ 750V  | 
 43.80 ~ 62.22 43.80 ~ 59.72  | 
 -  | 
 1.0(4)  | 
 -  | 
| 
 M & F  | 
 400mV 4V 40V 400V  | 
 20mV ~ 400mV 400mV ~ 4V 4V ~ 40V 40V ~ 400V  | 
 -31.76 ~ -5.74 -5.74 ~ 14.26 14.26 ~ 34.26 34.26 ~ 54.26  | 
 1.0 0.8 0.8 0.8  | 
 0.2 0.1 0.1 0.1  | 
 1.0 0.8 0.8 0.8  | 
| 
 1000V (dc) 750V (ac)  | 
 400V ~ 1000V 400V ~ 750V  | 
 54.26 ~ 62.22 54.26 ~ 59.72  | 
 -  | 
 1.0(4)  | 
 -  | 
| 
 (1) Auto-ranging is used when dBm function is selected (2) In Vac 750V range, 5% over-range is readable (3) Reading displayed in dB when REL modifier is used (4) For input voltage at frequency between 45Hz to 1kHz  | 
0dBm: 1 m Ohm @ 600 Ohm Reference Impedance Resolution: 0.01dB at slow and medium rate; 0.1dB at fast rate for all ranges. CMRR: > 90dB for dc signal Response Time: Same as ac voltage and ac current measurements. 
A-3 General Specifications
| 
 General Items  | 
 Specifications  | 
| 
  Ohmarm up time  | 
 At least 30 minutes  | 
| 
 Temperature Coefficient    | 
 Add 0.15 x (the applicable accuracy)/°C at 0°C to 18°C and 28°C to 50°C  | 
| 
 Operating Temperature  | 
 0°C to 50°C (32°F to 122°F)  | 
| 
 Storage Temperature  | 
 -20°C to 60°C  | 
| 
 Altitude  | 
 Up to 2000 M  | 
| 
 Pollution Degree  | 
 II  | 
| 
 Over-voltage Category  | 
 CAT II-600V and CAT I-1000V  | 
| 
 Relative Humidity        | 
 —  Up to 80% for 0°C to 28°C (75% RH for 12M Ohm and above ranges of resistance measurement) —  Up to 70%for 28°C to 35°C —  Up to 50%for 35°C to 50°C  | 
| 
 Common Mode Voltage    | 
 1000V dc or peak ac rms maximum between any input and earth ground  | 
| 
 Dimension  | 
 Approx. 255(w) x 105(h) x 305(d) mm (with holsters)  | 
| 
  Ohmeight  | 
 <3.0kgs  | 
| 
 Line Voltage  | 
 100V / 120V / 220V / 240V ac ±10%, 50/60Hz, 16VA maximum  | 
| 
 Interface              | 
 —  RS-232 (DB-9, male connector) —  Baud rates: 9600, 4800, 2400, 1200, 600, 300 —  Data length: 7 or 8 bits —  Parity: even / odd / none —  Stop bit: 1 or 2 bits —  Echo: on / off —  Print mode: on / off  | 
| 
 Safety Requirement  | 
 Designed in compliance with EN61010-1 (IEC1010-1)  | 
| 
 Installation Category    | 
 CAT-I 750VAC/1000VDC or CAT-II 600V,Pollution Degree 2 Environment  | 
| 
 EMC Requirement  | 
 Designed in compliance with EN61326-1.  | 
		
		 
		 定货信息:
		 
	     
		   
		     | 定货号
		      | 产品名称
		      | 规格配置
		      | 价格
		    | 
           | DA-01057-00 |  | 标准配置或电话咨询010-84851836 | 咨询 |